Resumen
ISO 5356-1:2004 specifies dimensional and gauging requirements for cones and sockets intended for connecting anaesthetic and respiratory equipment, e.g. in breathing systems, anaesthetic-gas scavenging systems and vaporizers.
ISO 5356-1:2004 gives requirements for the following conical connectors:
- 8,5 mm sizes intended for use in paediatric breathing systems;
- 15 mm and 22 mm sizes intended for general use in breathing systems;
- 22 mm latching connectors (including performance requirements);
- 23 mm size intended for use with vaporizers, but not for use in breathing systems;
- 30 mm size intended for the connection of a breathing system to an anaesthetic gas scavenging system.
ISO 5356-1:2004 does not specify the medical devices and accessories on which these connections are to be provided.
Requirements for the application of conical connectors are not included in ISO 5356-1:2004, but are or will be given in the relevant International Standards for specific medical devices and accessories.
NOTE Requirements for screw-threaded weight-bearing conical connectors are specified in ISO 5356-2.
Informaciones generales
-
Estado: RetiradaFecha de publicación: 2004-05Etapa: Retirada de la Norma Internacional [95.99]
-
Edición: 3Número de páginas: 16
-
Comité Técnico :ISO/TC 121/SC 1ICS :11.040.10
- RSS actualizaciones
Ciclo de vida
-
Anteriormente
RetiradaISO 5356-1:1996
-
Ahora
-
Revisada por
PublicadoISO 5356-1:2015