ISO 5356-1:2004 specifies dimensional and gauging requirements for cones and sockets intended for connecting anaesthetic and respiratory equipment, e.g. in breathing systems, anaesthetic-gas scavenging systems and vaporizers.
ISO 5356-1:2004 gives requirements for the following conical connectors:
- 8,5 mm sizes intended for use in paediatric breathing systems;
- 15 mm and 22 mm sizes intended for general use in breathing systems;
- 22 mm latching connectors (including performance requirements);
- 23 mm size intended for use with vaporizers, but not for use in breathing systems;
- 30 mm size intended for the connection of a breathing system to an anaesthetic gas scavenging system.
ISO 5356-1:2004 does not specify the medical devices and accessories on which these connections are to be provided.
Requirements for the application of conical connectors are not included in ISO 5356-1:2004, but are or will be given in the relevant International Standards for specific medical devices and accessories.
NOTE Requirements for screw-threaded weight-bearing conical connectors are specified in ISO 5356-2.
Общая информация
-
Текущий статус: ОтозваноДата публикации: 2004-05Этап: Отмена международного стандарта [95.99]
-
Версия: 3
-
Технический комитет :ISO/TC 121/SC 1ICS :11.040.10
- RSS обновления
Жизненный цикл
-
Ранее
ОтозваноISO 5356-1:1996
-
Сейчас
-
00
Предварительная стадия
-
10
Стадия, связанная с внесением предложения
-
20
Подготовительная стадия
-
30
Стадия, связанная с подготовкой проекта комитета
-
40
Стадия, связанная с рассмотрением проекта международного стандарта
-
50
Стадия, на которой осуществляется принятие стандарта
-
60
Стадия, на которой осуществляется публикация
-
90
Стадия пересмотра
-
95
Стадия, на которой осуществляется отмена стандарта
-
00
-
Пересмотрен
ОпубликованоISO 5356-1:2015